Ludwig,
Your question regarding the interest of using a B&D daily and to keep the annual validation for the sterilisers is very interesting for many reasons. As you said, sterilisers are now more and more well programmed with a lot of new technologies and monitoring.
In the most european countries, the autoclaves are validated once a year + each time after maintenance interventions or after a change of product, packaging etc... This is important to do it annualy to have a "reference" for each cycle with each typical load used in the CSSD.
Regarding the B&D test every morning we should actually speak in terms of "steam penetration test", that allows the CSSD manager to do it with different methods than the classic B&D sheet. There is now some electronic monitoring probes used for this but also for each cycle. These kind of systems are interesting because it is totally independant from the autoclave.
This is also a way to have a load validation system for each load, in each steriliser, based on physical measurement. Used in routine it can help to check daily if your autoclave is still under the last anual validation parameters and so, prevent any maintenance from the biomedical department or the manufacturer of the machines itself.
It exist an international standard which is from my point of view the only one to speak clearly about the load validation in routine:
The EN / ISO 14937 from year 2000:
Sterilization of health care products - General requirements for characterization of a sterilizing agent and the development, validation and routine control of a sterilization process for medical devices
Stérilisation des produits de santé - Exigences générales pour la caractérisation d'un agent stérilisant et pour le développement, la validation et la vérification de routine d'un processus de stérilisation pour dispositifs médicaux
10 Routine monitoring and control
10.1 The purpose of routine monitoring and control is to demonstrate that the validated and specified sterilization process has been delivered to the product.
10.2 There shall be evidence through measurements, supplemented as necessary by biological indicators or chemical indicators, that the sterilization process was delivered within the defined tolerances (see also 9.4.3).
10.3 Data shall be recorded to demonstrate the attainment of process parameters.
10.4 All records shall be retained in accordance with 4.2.4.
10.5 If biological indicators are used in routine monitoring, they shall comply with 8.3 a) and b).
10.6 If chemical indicators are used in routine monitoring, they shall comply with ISO 11140-1 and any subsequent parts of ISO 11140 that are applicable to the process.
Best Regards
Michael Fangon
|