Date: 24 September 2002 [GMT] Subject: Prevac Cycle vs. Gravity Cycle |
||
When a manufacturer give recommendations to sterilize an item in a prevac cycle, can you also run that item in a gravity cycle since the gravity cycle is a lower temperature? My thoughts are that you sterilize an item the way the manufacturer recommends because that is the way they validated sterilization. |
From: Wim Renders (Belgium) Date: 2 October 2002 [GMT] Subject: Re: Prevac Cycle vs. Gravity Cycle |
Dear Jamie,
Because of these problems: According to the ISO and CEN international standards on sterilization, the process with downward displacement is only allowed for unwrapped instruments and bottled fluids." |