EFHSS - Questions & Answers - Steam Sterilization - Q00242
Pressure ripples during Sterilisation holding time
Follow Up
Web Search
From: (Kuwait)   Date: 2 March 2003, 09:39 [GMT]
Subject: Pressure ripples during Sterilisation holding time

Dear Madam/Sir,

Would you please explain:
1) During Sterilization holding Time there appears to be pressure ripples instead of a flat straight line (During 7 mins holding time). It is noticed that the pressure remains at 2.21 Bar and the Sterilization time at 135.3 Degrees Celcius in this case. Could it be due to the enlarged bleeder hole or a bad steam trap? Please advice.

With Best regards,

Joe Colaco,
Kuwait

From: (Netherlands)   Date: 3 February 2003, 07:26 [GMT]
Subject: Re: Pressure ripples during Sterilisation holding time

Hello Joe,

It is difficult to judge about these kind of problems without knowing the previous situation. You should compare the graphs of your recorder with the ones made before you noticed this ripples. It is important to know the method used by the steriliser to control the pressure and temperature. This means the steriliser control and the valves in the P&ID which are actually executing the instructions coming from the control. If you could send me a copy of the original graph and the one with the problem, by fax, I will give it a try to evaluate it and advise you. However if the ripple is not causing a too low sterilising temperature there seems to be no problem at all. Of course if you want to stick to the EN 285 the temperature ripple, including the theoretical temperature, should not exceed 1,5 degree Celsius.
If you have further questions, please do not hesitate to contact me directly.

Best Regards,

Henk Ras
Export Manager

Sanamij B.V.- Rotterdam, The Netherlands
Phone + 31 10 286 1750
Fax + 31 10 456 5697
website: www.sanamij.com