EFHSS - European Forum for Hospital Sterile Supply :: Questions & Answers EFHSS Questions and Answers - Question Q00040 - English Version
  <<>>

EFHSS Questions and Answers - Question Q00040

Categories Authors Questions Unanswered Questions Questions with Answers Questions and Answers Submit New Question Help

Category Sort By Sort Order
View

EFHSS - Questions & Answers - Steam Sterilization - Q00040
Autoclave Temperature Recording
Follow Up
Web Search
From: (United States)   Date: 3 October 2000, 18:02 [GMT]
Subject: Autoclave Temperature Recording

When documenting the temp on the autoclave are we to record the highest or the lowest temp?

From:    Date: 4 October 2000, 09:34 [GMT]
Subject: Re: Autoclave Temperature Recording

Good day Marcy,

I don't know any regulations or recommendations for this. At my institution, I have my staff document the last temperature of the sterilization portion of the cycle while verifying that the entire sterilization portion of the cycle was above the minimum required temperature.
If I was chosing between using the highest or lowest temperature, I would chose the lowest temperature, as this will tell you if the entire sterilization portion of the cycle was above the minimum temperature selected.
Because the death rate of micoorganisms doubles with every 10o C increase in temperature, any increase of the temperature over the minimum required temperature will increase the effectivness of the sterilization process. Therefore, the lowest temperature will give you the most critical information in monitoring that sterilization parameters have been met.

Regards, Pete Bobb