EFHSS - Questions & Answers - Steam Sterilization - Q00471
Bowie Dick Test for Steam Sterilizer
Follow Up
Web Search
From: (India)   Date: 24 May 2004, 10:29 [GMT]
Subject: Bowie Dick Test for Steam Sterilizer

I would like to know whether HTM2010 Part -3, states that for Bowie Dick Test on 121 degree temperature, the time required is 19 minutes? Please let me have the details on this standard.

From: Miki   Date: 24 May 2004, 20:50 [GMT]
Subject: Re: Bowie Dick Test for Steam Sterilizer

Hi,

B&D Test comes to approve an efficiant air removal stage. A positive result will assure good penetration of the steam into porous load.
The configuration of this stage is not depended on exposure temperature. This stage shall be identical in every cycle INTENDED for porous load (121,126,134 & BD Test). The exposure period depends on the Holding temperature.
The HTM 2010-3 states (13.48) holding time of 16.8 to 17.0 minutes at 121C.

miki

Please use the Follow Up form if you want to post a reply or answer to question Q00471. For new questions please use the Submit New Question form.

  ^<>>