EFHSS  European Forum for Hospital Sterile Supply

EFHSS Questions and Answers - Answer to Question Q00049

Categories Authors Questions Unanswered Questions Questions with Answers Questions and Answers Submit New Question Help

Category Sort By Sort Order
View

Please fill out the following form and submit it to post a reply and/or answer to question Q00049 to the EFHSS Questions and Answers pages. For new questions please use the Submit New Question form instead.

EFHSS Questions and Answers - Answer to Question Q00049
Your Details:
Name:
Email:
Country:
Your answer/reply:
Message:
 

Please contact/notify if you have any questions and/or problems using the EFHSS Questions and Answers pages.

EFHSS - Questions & Answers - Cleaning & Disinfection - Q00049
Decontamination and Cleaning of Theatre Instrument Trolleys
Follow Up
Web Search
From: (New Zealand)   Date: 22 December 2000, 02:59 [GMT]
Subject: Decontamination and Cleaning of Theatre Instrument Trolleys

Hi

I am a Team Leader for Quality and Training in the "Sterile Production Centre" at Wellington Hospital.

Can anyone help me with "standards and procedures" for the decontamination and cleaning of theatre instrument trolleys.

At the present all trolleys come to our department with the drapes and dirty instruments on. We are currently washing, hosing then drying.

Regards
Judy Hatton

From: Josy Holdener (Switzerland)   Date: 26 December 2000, 09:34 [GMT]
Subject: Re: Decontamination and Cleaning of Theatre Instrument Trolleys

Dear Judy,

"Standards and Procedures" connected with transportation of used Materials to Central Supply Dep. are e.g.

  • Devise should be cleaned of gross debris at the point of use. Textiles drapes should be placed in laundry bags and returned to the laundry service (not CSD), paper drapes discarded at the point of use.
  • Contaminated (all used) items should be contained during transport from the point of use to the decontaminated area. Bins with lids, enclosed or covered carts, closed sterilization containers, may be used alone or in combination to transport contaminated items.
  • Carts and other mobile equipment must be cleaned either in an automatic cart washer or by hand prior to reuse for sterile supplies of returning to the OR or patient care units.

References:

  • "Australian Standards"; Code of practice for cleaning, disinfecting and sterilizing reusable medical and surgical instruments and equipment, and maintenance of associated environments in health care facilities
    Published by Standards Australia,
    Sale: 1 The Crescent, Homebush , NSW. Tel. (02) 746 4600
  • Nat. Standards and Recommended Practices for Sterilization
    published by the Association for the Advancement of Medical Instrumentation
    3330 Washington Boulevard, Arlington, Virginia 22201, USA

Kind regards
Josy Holdener