EFHSS - Questions & Answers - Cleaning & Disinfection - Q00629
Compatibility of EAW
Follow Up
Web Search
From: (Thailand)   Date: 8 October 2004, 09:10 [GMT]
Subject: Compatibility of EAW

Hello

I need to know that endoscope from wich manufacturer (Olympus, Pentax, Fuginon) confirmed to disinfect by electrolyted acid water (EAW). My department plan to replace glutaraldehyde with EAW.

Thank you!!!

Nithat Sirij
Paolo Memorial Hospital

From: (Belgium)   Date: 11 October 2004, 10:54 [GMT]
Subject: Re: Compatibility of EAW

Dear,

preliminary test with EAW have pointed out an high degree of incompatibility with insertion tubes of endoscopes. Final test results will be available in coming months

Kind regards
Dr Ives Verheyleweghen
General Manager
Olympus Belgium

From: (Thailand)   Date: 19 October 2004, 06:23 [GMT]
Subject: Re: Compatibility of EAW

Dear Sir,

Could you tell me more the test about the process, what EAW system (I know CLEANTOP & Sterilox) and result.

Regards,
Nithat Sirij

From: (Germany)   Date: 27 October 2004, 18:58 [GMT]
Subject: Re: Compatibility of EAW

Dear Mr. Mrs. Sirij,

to answer your question, the company Fujinon issued already in the Year 2001 a full compatibility approval to disinfect Fujinon endoscopes in EAW generated by the Cleantop System. In adddition are PMS studies and experience reports on compatibility of flexible endoscopes available. For further information please contact

CBC (Deutschland) GmbH
Hansaalle 191 40549 Düsseldorf Germany
Email for direct communication:

Kind regards

  ^<>>
EFHSS - European Forum for Hospital Sterile Supply :: Questions & Answers EFHSS Questions and Answers - Answer to Question Q00629 - English Version

© 2005 EFHSS " Updated: 21 April 2005, 10:27 [GMT]