Dry Heat Sterilization
Follow Up
Web Search
From: (Iran)   Date: 13 March 2005, 14:20 [GMT]
Subject: Dry Heat Sterilization

Dear Sir/Madam

It would be appreciated if you could kindly advise me about Biological Test for dry heat sterilizer.

From: Josy Holdener (Switzerland)   Date: 13 March 2005, 17:07 [GMT]
Subject: Re: Dry Heat Sterilization

Dear
The spore used for testing dry-heat sterilizers is Bacillus atrophaeus(formerly known as Bacilllus subtilis). It is recommended to run a spore test with each load.
Kind regards
Josy Holdener

[15/03/2005] ...submittedFrom: Peter Newson (United Kingdom)   Date: 15 March 2005, 14:19 [GMT]
Subject: Re: Dry Heat Sterilization

Dear Rouhi

I agree with the answer from Josy - however the use of a biological indicator in every load is a matter for debate - as is the case with steam sterilization - in many countries around the world current practices include the use of biological indicators as part of your validation and re-validation processes but does not form part of your load control - the inclusion of a good quality validated chemical indicator is prefered.

Regards
Peter

Please use the Follow Up form if you want to post a reply or answer to question Q00764. For new questions please use the Submit New Question form.