EFHSS - European Forum for Hospital Sterile Supply

visit the EFHSS Industrial Partners...

Developed by Baumeister Mediasoft Engineering

EFHSS - European Forum for Hospital Sterile Supply

EFHSS Questions and Answers - Answer to Question Q01142 - English Version

HomeNewsSearchSitemapContact

EFHSS - European Forum for Hospital Sterile Supply :: Questions & Answers

EFHSS Questions and Answers - Answer to Question Q01142

Categories Authors Questions Unanswered Questions Questions with Answers Questions and Answers Submit New Question Help

Category Sort By Sort Order
View

We would like to remind you that this site's primary aim is to provide information to those who are interested in the reprocessing of medical devices in a hospital. We are not a forum for industrial sterilization as we do not have the necessary know how about this area. Can we ask you to respect the criteria of scientific objectivity when answering questions.

Thank you for your co-operation.

Please fill out the following form and submit it to post a reply and/or answer to question Q01142 to the EFHSS Questions and Answers pages. For new questions please use the Submit New Question form instead.

EFHSS Questions and Answers - Answer to Question Q01142
Your Details:
Name:
Email:
Country:
Your answer/reply:
Message:
 

Please contact/notify if you have any questions and/or problems using the EFHSS Questions and Answers pages.

EFHSS - Questions & Answers - Steam Sterilization - Q01142
Bowei & Dick test
Follow Up
Web Search
From: (Kuwait)   Date: 11 March 2006, 09:07 [GMT]
Subject: Bowei & Dick test

Dear Sir/Madam

  1. Regard Bowie & dick program in the benchtop autoclave, there are some autoclaves which do not have this program BUT have HELIX TEST program instead, can we use this program for B&D test packets?
  2. If the benchtop autoclaves does not have both B&D test program neither HElix test program, is it possible to use wrapped program (full cycle) to perform B&D test?

plz answer A.S.A.P (very urgent)
Dr, ghaneema Al-dakheel
superintendent of quality assurance & infection control in dental service

From: miki   Date: 12 March 2006, 18:34 [GMT]
Subject: Re: Bowei & Dick test

Dear Dr.
Both B&D AND Helix challenge the air removal from the chamber. B&D Test confirm effective air removal from porous load (fabric, wrapped or packed instruments).
Helix Test confirm successful air removal from "Hollow A" instruments like dental Hand pieces (Turbines), or in other words - capilars.
Both are tested at "BD Test" program.

BD and Helix tests are performed only on Autoclaves equipped with forced air removal (pump, ejector). Yes. take a 134C / 4minutes / 1-2 drying. But again: only if it is equipped with "pre-vacuum" stage.