EFHSS - European Forum for Hospital Sterile Supply

visit the EFHSS Industrial Partners...

Developed by Baumeister Mediasoft Engineering

EFHSS - European Forum for Hospital Sterile Supply

EFHSS Questions and Answers - Answer to Question Q00977 - English Version

HomeNewsSearchSitemapContact

EFHSS - European Forum for Hospital Sterile Supply :: Questions & Answers

EFHSS Questions and Answers - Answer to Question Q00977

Categories Authors Questions Unanswered Questions Questions with Answers Questions and Answers Submit New Question Help

Category Sort By Sort Order
View

We would like to remind you that this site's primary aim is to provide information to those who are interested in the reprocessing of medical devices in a hospital. We are not a forum for industrial sterilization as we do not have the necessary know how about this area. Can we ask you to respect the criteria of scientific objectivity when answering questions.

Thank you for your co-operation.

Please fill out the following form and submit it to post a reply and/or answer to question Q00977 to the EFHSS Questions and Answers pages. For new questions please use the Submit New Question form instead.

EFHSS Questions and Answers - Answer to Question Q00977
Your Details:
Name:
Email:
Country:
Your answer/reply:
Message:
 

Please contact/notify if you have any questions and/or problems using the EFHSS Questions and Answers pages.

EFHSS - Questions & Answers - Cleaning & Disinfection - Q00977
Stick on Glut or change to OPA?
Follow Up
Web Search
From: (Malaysia)   Date: 26 September 2005, 08:38 [GMT]
Subject: Stick on Glut or change to OPA?

Dear All,

We've been using glutaraldehyde (GA) since many years ago. And off and on we have heard of complain about the "standard" GA side effects (headache, irritation etc). We've heard that OPA is a good substitute to GA. But, We have found that some of the OPA users experienced the similar kind of side effects!

In view that we are a small setting with small budget, we feel that GA still be the most economical HLD.

Please kindly advise.

Thanks.

From: (Israel)   Date: 29 September 2005, 07:38 [GMT]
Subject: Re: Stick on Glut or change to OPA?

Dear Leon,
Because OPA is relatively new, there are only a few references and evidence about its potential health effects. Like glutaraldehyde, OPA may cause allergic reactions and occupational asthma; however the potential for airborne exposure appears to be reduced. Regardless of which chemical you choose, all high-level disinfectants should be treated with caution. OPA is probably a sensitizer and should be treated as such. Lutaraldehyde and OPA should be used with local exhaust ventilation; staff should use personal protection such as impermeable gloves and gowns to prevent any dermal exposure from handling these products.

Regards,
Yaffa Raz

From: (Belgium)   Date: 29 September 2005, 14:34 [GMT]
Subject: Re: Stick on Glut or change to OPA?

Regarding GA / OPA (or also PA: peracetic acid): as manufacturer of both chemicals we seen in europe a trend away from GA towards desinfection in machines (also using more and more PA - see Olympus scope desinfectors) and towards PA as manual desinfection

we do not see the same effects on health care workers as we see with GA

with PA: be carefull with "old" instruments and always ask the certificates from your supplier concerning the use of a PA on your scopes/instruments

hope this helps

Marc.

From: Leon   Date: 30 September 2005, 04:22 [GMT]
Subject: Re: Stick on Glut or change to OPA?

Thank you for the advise!

From: (Belgium)   Date: 2 February 2006, 14:07 [GMT]
Subject: Re: Stick on Glut or change to OPA?

Dear,

The instructions for use of Cidex OPA say also the following under contraindications:

  1. 1. CIDEX OPA Solution should not be utilized to process any urological instrumentation used to treat patients with a history of bladder cancer. In rare instances CIDEX OPA Solution has been associated with anaphylaxis-like reactions in bladder cancer patients undergoing repeated cystoscopies.
  2. 2. CIDEX OPA Solution should not be utilized to process instrumentation for patients with known sensitivity to CIDEX OPA Solution or any of its components.
  3. 3. CIDEX OPA should not be used to sterilize heat sensitive medical devices.

I hope this information still can be helpful to you.
Best regards,
Wim Renders